VCP6xPN Vacuum Transducer ● Profinet ● Pirani ● 1000 to 5e-4 mbar

 648.00 697.00 excl. VAT and transport costs

The processor controlled VCP Pirani vacuum transducer with robust, elastic helix filament is ideal for precise measurements in demanding applications with aggressive media.

Clear
Compare

Additional information

Weight1.1 kg
Dimensions25 x 16 x 9 cm
Measurement principle

Gas dependency

depending on gas type

Measurement range mbar

Measurement range Torr

750 – 5e-4 Torr

max. overload

10 bar abs., 16 bar abs. with CERT31P

max. overload with CERT31P

up to 16 bar abs.

Accuracy

1000 – 10 mbar: ca. 30 % f. r. | 10 – 0.01 mbar: 10 % f. r.

Repeatability

10 – 0.01 mbar: 5 % f. r.

Materials in contact with vacuum

Stainl. steel 1.4307, platinum /rhodium, nickel, glass

Response time

40 ms

Operating temperature

+5…+50°C (Profinet)

Storage temperature

-40…+65 °C

Bake out temperature

Voltage supply

20 – 30 VDC

Power consumption

Max. 2.5 W, add. 0.8 W f. EtherCAT/relays/ LCD, add. 1.6 W for Profinet

Serial interface

Electrical connection

Article Number (choice of vacuum connection):

,

Dimensions

99 x 69 x 48 mm (VCP63)

Protection class

Weight

190 g (VSP63)

Further relevant products